Anesthesia Workstation - 300
The ADS-300 is a sophisticated anesthesia delivery system featuring an integrated microprocessor-based anesthesia ventilator and gas administrator. Designed for user-friendly operation, the ADS-300 is suitable for both adult and pediatric patients. Its 7-inch TFT touchscreen display provides clear visuals of flow, volume, and pressure graphs.
The integrated ventilator incorporates electric-driven bellows, eliminating the need for driving gas. Additionally, a lithium-ion battery ensures reliable operation for up to 2 hours. The ADS-300 is equipped with a high-accuracy 3-gas, 5-tube rotameter with an anti-hypoxic device and a standard Select-a-Tek bar for twin vaporizers.
To enhance safety, the system incorporates Fresh Gas Compensation and Fresh Gas Decoupling features. These functionalities contribute to more reliable and secure anesthesia delivery.
Prime Features
> Cylinder and Pipeline Pressure Monitoring Gauges
> High accuracy 3 Gas 5 Tube rotameter with anti-hypoxic device
> Compact rotary switches for ON/OFF, N20/air, and
ACGO/Closed
> Offering ventilation modes including VCMV, V-SIMV, PCMV, P-SIMV, PRVC, PSV, BAG, and
standby
> Features a user-friendly touchscreen interface
> Quick and easy BAG/VENT changeover
> Automated canister bypass for in-surgery refilling
> Eliminated complications with integrated flow sensor
> Air filter for continued operation in atmospheric air, even during total fresh gas failure
> NIST-standard inlet
ports for oxygen, N20, and air
> Equipped with an additional power outlets
> Standard ACGO connection
> Fresh gas decoupled emergency oxygen flush
> Standard index-pinned cylinder yokes for oxygen and N20
> Equipped with an
auxiliary DIN standard
oxygen outlet
Specification
Control Parameters | Range/Details |
---|---|
Mode of Operation | VCMV, V-SIMV, PCMV, P-SIMV, PRVC, PSV, BAG, Standby |
Patient Type | Adult and Child (Selectable) |
Tidal Volume | 50 to 1500 ml |
Ti (Inspiration Time) | 0.2 to 8.0 Sec |
Pi (Inspiratory Pressure) | 5 to 60 cmH₂O |
Respiratory Rate | 6 to 60 Bpm |
Plateau | 0 to 50% |
PEEP | 0 to 30 cmH₂O |
Pressure Support | 5 to 40 cmH₂O |
Trigger | 0.5 to 5.0 cmH₂O |
Backup RR | 6 to 60 Bpm |
Backup Mode | VCMV, V-SIMV, PCMV, P-SIMV, PRVC |
Category | Details/Range |
---|---|
Alarms and Indications | |
Inspiratory Pressure High | 10 to 60 cmH₂O |
Inspiratory Pressure Low | 5 to 15 cmH₂O |
Minute Volume High | 3 to 50 Liters |
Minute Volume Low | 0.1 to 47 Liters |
RR High | 15 to 100 Bpm |
RR Low | Specified alarm for low respiratory rate |
FIO₂ Alarms | Specified alarm for oxygen fraction levels |
Monitoring Parameters | Tidal Volume, Minute Volume, RR, FIO₂, PI, PEEP Obtained |
Loop Types | Flow vs Pressure, Flow vs Volume, Volume vs Pressure (Selectable Any Two) |
Graph Types | Flow, Pressure, Volume (Selectable Any Two) |
General Parameters | Details/Range |
---|---|
Display | 7″ TFT with Touch Screen |
Power Input | 90 to 260 V AC, 50 Watts, 50/60 Hz |
Pipe Line Inlet (O₂, N₂O & Air) | 2.4 – 6 Bar |
Cylinder Inlet (O₂ & N₂O) | 150 Bar Max |
Battery Backup | 2 hours (Lithium-ion) |
Respiratory Rate (RR) | 5 to 90 Bpm |
Oxygen Concentration (FIO₂) | 21% to 100% |
Alarms | Low and High for Pressure, RR, Minute Volume, and FIO₂; Active patient; Patient disconnected; Oxygen Pressure Low; Low battery indication; Apnea |