Metsl Vital Track 12

  • 12.1″ Color LCD with multichannel waveform display.
  • Light Weight, Easy to carry, suitable for ambulance center
  • Easy connecting with central monitoring system via Ethernet/Wireless
  • 200 Hours Data Storage
  • Simple & Easy operation
  • Suitable for adult, pediatrics and neonates
  • Standard parameters: 3/5/12 Lead ECG, SPO2, Nibp, 2-Temp
  • Optional parameters:- EtCO2, 2 IBP
  • Thermal Recorder & wall installation
  • Controlled by both knob and touch screen.

Technical Specification

ECG

ParameterDetails
Electrode3/5/12 Leads
Numeric Data Display6 lead ECG with HR, Heart Rate, VPC rate (per min.), ST Level, respiration rate, NIBP (Systolic, Diastolic, MAP), SpO2 (Nellcor), Pulse rate, Temperature, Heart rate sync mark, Pulse rate sync mark, respiratory sync mark
LeadLaVR, aVL, aVF, V (5 Lead), III or III (3 Lead), JJIIlaVR, avi, aVF, V1-V6 (12 Lead)
Gain Selectionx1/4, x1/2, x1 and x2
Frequency ResponseDiagnostic: 0.05 to 90 Hz, Monitor: 0.5 to 35 Hz, Surgery: 1-15 Hz
Calibration Signal1 (mV p-p), Accuracy: ±5%
AnalysisST Segment & Arrhythmia Detection
Display Channel6 Channels
Sweep Speed6.25, 12.5, 25 or 50 mm/s

Nibp

ParameterDetails
TechniqueOscillometric during inflation
RangeAdult: 0-300 mm Hg
Pediatric: 0-250 mm Hg
Neonate: 0-135 mm Hg
Measurement Cycle<40 sec, typical
Cycles (Selectable)2, 3, 5, 10, 15, 30 min, 1, 2, 4 hr
Max. Allowable Cuff PressureAdult: 297 ±3 mm Hg
Pediatric: 240 ±3 mm Hg
Neonate: 147 ±3 mm Hg
Resolution1 mm Hg
Transducer Accuracy+2 mm Hg or +2% of reading
ModesAuto, Manual & STAT

Pulse Oximetry (Nellcor)

ParameterDetails
Range0 to 100%
Resolution1%
Accuracy70% to 99% range ±1%, 40 to 69% range ±1%
MethodDual wavelength LED

Heart (Pulse) Rate

ParameterDetails
SourceUser Selectable: Smart, ECG, Plenth, Nibp
RangeAdult: 15-300 bpm
Pediatric: 15-350 bpm
Accuracy+1 bpm or ± 1%

Environment

ParameterDetails
Operating Temperature10-40 °C
Storage Temperature-5-50 °C
Humidity15% to 90%

Temperature (Dual Channel)

ParameterDetails
Range, Accuracy0-50 °C ±1%
Display Resolution±0.1°C
ProbeYSI 400 and YSI 700 Series
Probe Gradation1/10 Degree

IBP (Optional)

ParameterDetails
Channels2 Nos
PressureART, PA, CVP, RAP, LAP, ICP P1, P2
Range0-300, 6-120, 10-300 mmHg
Accuracy+1 mmHg or ± 2%

Trends

ParameterDetails
Memory200 Hr at 30 seconds intervals for a NIBP reading every 5 minutes
DisplayTabular, graphical

General

ParameterDetails
Dimensions320 (L) x 280 (B) x 135 (T) mm
Weight3.5 Kg
PowerAC 100-240 V, 50/60 Hz
Power Consumption35 W
Battery14.8 V/2200 Mah Li-Ion
Recharge Time4.5 hours
Operating Time> 2 Hours typical
Alarm Silence120 seconds
Alarm Type3 Level adjustable Audio-visual alarms for each parameter, Sensor/Cuff disconnect, Low battery
Electrical SafetyESU, Defibrillator protection
ConnectivityWiFi/ USB/Ethernet port/ LAN

EtCO2 (Optional)

ParameterDetails
Transducer TypeSide stream CO2 sensor
Range0-150 mm Hg
Resolution0.1 mm Hg
Accuracy0-40 mmHg: ±2 mm
41-70 mmHg: ±5% of reading
71-100 mmHg: ±8% of reading
101-150 mmHg: ±10% of reading

Respiration Rate

ParameterDetails
MethodImpedance Pneumography
RateAdult: 0-120 rpm
Neo/Ped: 0-150 rpm
Threshold ControlYes
Resolution1 rpm
Accuracy±2 breath/min